Specialist for hygienic applications, as remote version with up to 4 I/Os

Specs at a Glance

Parameter Specification
Measurement Error Volume Flow (Standard): ±0.5 % o.r. ± 1 mm/s (0.04 in/s)
Volume Flow (Option): ±0.2 % o.r. ± 2 mm/s (0.08 in/s)
Measuring Range 0.06 dm³/min to 600 m³/h (0.015 gal/min to 2,650 gal/min)
Medium Temperature Range –20 to +150 °C (–4 to +302 °F)
Process Pressure PN 40, Class 150, 20K
Wetted Materials Liner: PFA
Electrodes: 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); Tantalum; Platinum
Process Connections: Stainless Steel 1.4404 (316L); PVDF; PVC Adhesive Sleeve
Seals: O-ring Seal (EPDM, FKM, Kalrez); Aseptic Molded Seal (EPDM, FKM, Silicone)
Grounding Rings: Stainless Steel 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); Tantalum

Description

See all specifications

Field of application

Promag H is the preferred sensor for hygienic applications with highest requirements in the food and beverage and life sciences industries. With its innovative remote transmitter Promag H 500 maximizes installation flexibility and operational safety in demanding environments. Heartbeat Technology enables compliance and process safety at all times.

 

Benefits

  • Multivariable measurement for flow, temperature and conductivity
  • Flexible installation – numerous hygienic process connections and 0 x DN inlet run with constant accuracy independent of flow profile
  • Energy-saving flow measurement – no pressure loss due to cross-section constriction
  • Maintenance-free – no moving parts
  • Full access to process and diagnostic information – numerous, freely combinable I/Os and Ethernet
  • Reduced complexity and variety – freely configurable I/O functionality
  • Integrated verification – Heartbeat Technology
  • The bidirectional measuring principle is virtually independent of pressure, density, temperature and viscosity
  • Dedicated to demanding applications in the food and beverage as well as in life sciences industries


Device properties

  • Liner made of PFA
  • Sensor housing made of stainless steel (3-A, EHEDG)
  • Wetted materials CIP, SIP cleanable
  • Remote version with up to 4 I/Os; hygienic sensor connection housing with IP69
  • Backlit display with touch control & WLAN access
  • Standard cable between sensor & transmitter
  • Measuring principle

Electromagnetic

  • Product headline

Specialist for hygienic applications, as remote version with up to 4 I/Os.
Dedicated to demanding applications in the food and beverage as well as in life sciences industries.

  • Sensor features

Flexible installation concept – numerous hygienic process connections. Energy‐saving flow measurement – no pressure loss due to cross section constriction. Maintenance‐free – no moving parts.
Liner made of PFA. Sensor housing made of stainless steel (3-A, EHEDG). Wetted materials CIP, SIP cleanable.

  • Transmitter features

Full access to process and diagnostic information – numerous, freely combinable I/Os and fieldbuses. Reduced complexity and variety – freely configurable I/O functionality. Integrated verification – Heartbeat Technology.

  • Nominal diameter range

DN 2 to 150 (¹⁄₁₂ to 6″)

  • Wetted materials

Liner: PFA
Electrodes: 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); Tantalum; Platinum
Process Connections: stainless steel, 1.4404 (F316L); PVDF; PVC adhesive sleeve
Seals: O-ring seal (EPDM, FKM, Kalrez), aseptic molded seal (EPDM, FKM, silicone)
Grounding Rings: stainless steel, 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); tantalum

  • Measured variables

Volume flow, temperature, conductivity, mass flow, corrected volume flow, corrected conductivity

  • measurement error

Volume flow (standard): ±0.5 % o.r.± 1 mm/s (0.04 in/s)
Volume flow (option) ±0.2 % o.r. ± 2 mm/s (0.08 in/s)

  • Measuring range

0.06 dm³/min to 600 m³/h (0.015 gal/min to 2 650 gal/min)

  • process pressure

PN 40, Class 150, 20K

  • Medium temperature range

–20 to +150 °C (–4 to +302 °F)

  • Ambient temperature range

–40 to +60 °C (–40 to +140 °F)

  • Sensor housing material

1.4301 (304), corrosion resistant
Sensor connection housing (standard): AlSi10Mg, coated
Sensor connection housing (option): 1.4301 (304); 1.4409 (CF3M), similar to 316L

  • Transmitter housing material

AlSi10Mg, coated; 1.4409 (CF3M) similar to 316L; Polycarbonat

  • Degree of protection

Sensor remote version (standard): IP66/67, type 4X enclosure
Sensor remote version (option): IP69. Transmitter remote version: IP66/67, Type 4X enclosure

  • Display/Operation

4-line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible

  • Outputs

4 outputs:
4-20 mA HART (active/passive)
4-20 mA (active/passive)
Pulse/frequency/switch output (active/passive)
Relay output

  • Inputs

Status input
4-20 mA input

  • Digital communication

HART, PROFIBUS DP, PROFIBUS PA, FOUNDATION Fieldbus, Modbus RS485, Modbus TCP over Ethernet-APL, PROFINET, PROFINET over Ethernet-APL, Ethernet/IP, OPC-UA

  • Power supply

DC 24 V
AC 100 to 230 V
AC 100 to 230 V / DC 24 V (non-hazardous area)

  • Hazardous area approvals

ATEX, IECEx, cCSAus, INMETRO, NEPSI, EAC; JPN, UK Ex, KC

  • Product safety

CE, C-tick, EAC marking

  • Functional safety

Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511

  • Metrological approvals and certificates

Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)

  • Marine approvals and certificates

LR approval, DNV approval, ABS approval, BV approval

  • Pressure approvals and certificates

PED, CRN

  • Material certificates

3.1 material

  • Hygienic approvals and certificates

EHEDG, 3-A, liner and seals acc. to FDA, cGMP